Frequency Standard Catalog

 

HP 5071A Primary Frequency Standard

 

The HP 5071A Primary Frequency Standard delivers unsurpassed accuracy and stability for the most demanding laboratory and field applications. It is the first cesium standard to specify its stability for averaging times longer than a day. With a specification for 30-day averaging time, the HP 5071A's performance will ensure that your calibration services are available longer.

The HP 5071A is easy to use. Automatic startup is simple and requires no adjustments. A logical menu structure simplifies front-panel operations, selections, and status reporting. The HP 5071A can be operated and maintained anywhere. All controls are programmable, status can be checked remotely, and no adjustments or alignments are necessary during operation. An internal battery provides 45 minutes of backup in case of ac power failure.

Accuracy (full environmental range)
      
Temperature: 0° C to 50° C
Humidity: 0% to 80% (40° C maximum)
DC magnetic field: 0 gauss to 2 gauss
 
Accuracy:  ±2 X 10e-12 (±1 X 10e-12, Option 001)
Reproducibility:  ±5.0 X 10e-13
 
Warmup Time (typical): 15 min to normal operating status
                       30 min to full specifications
 
Settability: Resolution: 6.3 X 10e-15
             Range: ±1.0 X 10e-9
 
Frequency change due to environment: ±1 X 10e-13 (±8 X 10e-14, Option 001)
 
 
Stability (full environmental range)
 
Time domain stability (excluding environmental effects), 5/10-MHz outputs:
 
      Averaging Times
                        HP 5071A                HP 5071A Opt 001
      10e4 s            <= 5.6 X 10e-13         <= 8.5 X 10e-14
      1 day             <= 2.0 X 10e-13         <= 3.0 X 10e-14
      5 days            <= 1.0 X 10e-13         <= 2.0 X 10e-14
      30 days           <= 5.0 X 10e-14         <= 2.0 X 10e-14
 
 
Flicker floor (value)
<= 5.0 X 10e-14
<= 2.0 X 10e-14 (Option 001)
 
 
Frequency Domain Stability:
   
                                    SSB Phase Noise dBc
   
                                                High-performance
    Offset from         Standard cesium         cesium beam tube
    signal (Hz)            beam tube            (Option 001)
   




10e0 <= -85 <= -100 10e1 <= -125 <= -130 10e2 <= -135 <= -135 10e3 <= -140 <= -140 10e4 <= -145 <= -145 10e5 <= -145 <= -145     Sinusoidal output characteristics (all located on rear panel)   1 MHz Parameter Ports 1 and 2* 100 kHz
Amplitude into 50 ohm load >1 Vrms >1 Vrms   Isolation between ports >110 dB (typ) not specified   Harmonic and sub-harmonic signals (typical) <-40 dbc>Internal standby battery (nominal values)   Capacity: 45 minimum at 25° C from full charge Charge Time: 16 hour maximum from fully discharged state Charge Source: AC input power only   Remote system interface RS-232-C (DTE Configuration): Complete remote control of all instrument functions and parameters Connector: 9-pin male rectangular D subminiature type   Environmental and physical   Operating Temperature: 0° to 55°C Non-operating Temperature -40° to +70°C Size: 425.5 mm W X 133.4 mm H X 523.9 mm D (16.75 in X 5.25 in X 20.63 in) Weight: 30 kg (65 lb)   HP 5071A Primary Frequency Standard   Opt 001 High-Performance Cesium-Beam Tube Opt 0B2 Extra Operating and Programming Manuals Opt 0BW Extra Assembly-Level Service Manual Opt 0BV Component-Level Service Information Package Opt 908 Rack Flange Kit (for using the HP 5071A without its handles) Opt 913 Rack Flange Kit (for using the HP 5071A with its handles) Opt AX9 Rackmount Slides, Tilt Opt W30 Extended Repair Service, 2 years Opt W50 Extended Repair Service, 4 years (Opt W50 is not available with Opt 001)   Telecommunications Options   Opt 048 48 Vdc Power Opt 104 or 105 1.544 Mb/s, 100 ohm Balanced Opt 220, 221, or 222 2.048 Mb/s, 120 ohm Balanced Opt 270, 271, or 272 2.048 Mb/s 75 ohm     HP 5089A Standby Power Supply (approximately 8 hours standby time) HP 5087A Distribution Amplifier HP K34-59991A Phase Comparator PN 05089-60101 External dc power cable for standby power supply PN 1494-0059 Fixed rack slides PN 1494-0063 Tilt rack slides


 

1999.12.15. 263.4  15¡É
2000.01.15. 266.5  30¡É
2000.01.16  268.0  30¡É
2000.01.17  270.0  31¡É
2000.02.23  267.5  32¡É
2000.08.07  274.0  3?¡É
2000.08.20  274.0  3?¡É
2000.12.17  264.0  29.5¡É

.
.
.

2001. 9.15  270.0  28¡É

 

HP 105B Quartz Frequency Standard

 


The HP 105B is an economical, precision frequency standard with performance comparable to the best of the expensive primary standards. It is ideal for use in mobile and fixed installations, including applications in navigation and advanced communications. The 5-MHz output of the HP 105B has excellent short-term stability and spectral purity with a signal-to-noise ratio of better than 90 dB.

The HP 105B is based on the HP 10811D, an oven-controlled crystal oscillator with exceptional reliability. Its low aging rate and fast warm-up time reduce maintenance costs and downtime. Low power consumption extends battery backup time.


Frequency Stability:

 
        Long-Term Stability (Aging Rate): <5 x>Short-Term Stability: For 5 MHz output only. Refer to tables below.
 
 
                Table 1: Time Domain Stability
                -----------------------------------------
                Averaging time          Stability
                tau[seconds]            phi y (tau)
                -----------------------------------------
 
                10e-3           7.5 X 10e-10
                10e-2           1.5 X 10e-10
                10e-1           1.5 X 10e-11
                10e0            1 X 10e-11
                10e1            1 X 10e-11
                -----------------------------------------
 
 
 
                Table 2: Frequency Domain Stability
                -----------------------------------------
                Offset from 5 MHz       Phase Noise Ratio
                f [Hz]                  L(f) [dBc]
                -----------------------------------------
                10e0                    -90
                10e1                    -120
                10e2                    -126
                10e3                    -146
                10e4                    -150
                -----------------------------------------
 
        As a Function of Ambient Temperature: Frequency change is less than 
                2.5 X 10e-9 total from 0 degrees C to +50 degrees C.
 
        As a Function of Load: +/-2 X 10e-11 from open circuit to short, 
                50 ohm R, L, or C load change.
 
        As a Function of Supply Voltage: +/-5 X 10e-11 for 22-30V dc from 
                26V dc reference (battery operation, 105B) and for 115/230V +/-10%.
 
Frequency Adjustments:
 
        Fine Adjustments: Front-panel control with +/-5 X 10e-8 range, with 
                digital dial reading parts in 10e10.
 
        Coarse Adjustments: 1 X 10e-6, coarse control is screwdriver 
                adjustment at front panel.
 
Phase Locking Capability:
 
        An external +5V to -5V applied to EFC connector allows >2 parts in 10e8 
                frequency control for locking to an external source.
 
Warm-up (at 25 degrees C):
 
        To within 1 X 10e-7 in 15 min., 1 X 10e-8 in 20 min., 1 X 10e-9 in 30 min. 
                of final value (24 hours after turn-on).
 
Outputs:
 
        Frequencies: 5 MHz, 1 MHz, 100 kHz
 
        Voltage Levels: >=1V rms into 50 ohm; clock drive >=0.5V rms into 
                1 Mohm.
 
        Connectors: 5 MHz, 1 MHz, 100 kHz front- and rear-panel BNC 
                connectors, clock drive, and electronic frequency control 
                rear-panel BNC connectors.
 
        Harmonic Distortion: (5 MHz, 1 MHz,  100 kHz) down more than 
                40 dB from rated output.
 
        Spurious Phase Modulation: (Discrete sidebands 10 Hz to 25 kHz) 
                down more than 80 dB from rated output (5 MHz, 1 MHz, 100 kHz).
 
 
HP 105B Quartz Frequency Standard
 
        Opt 908 Rack Flange Kit
 
 
PN 5060-0049    Board Extender, 15 pin
PN 5060-0630    Board Extender, 22 pin
PN 1251-0126    Connector (EXT DC, 5 pin)
 
HP 5085A        Standby Power Supply (40 to 70 hours)
 

 

°ü·Ã Site

 

Çѱ¹Ç¥ÁذúÇבּ¸¿ø - ±¹°¡ÁöÁ¤½Ã°£Á֯ļö ¿¬±¸½Ç¿¬±¸½Ç

General Time and Frequency Topics


National Laboratories / Standards Organizations


Time Transfer Signals